6–2–13 Monitor15 Cover Set (FRU No. 171)
Time Required
20 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6–25 on page 6–87.
1.
Turn OFF the system.
Unscrew two screws ( 1 – 2 ).
2.
3.
Remove the LAMP SHADE2 (FRU179).
Unscrew six screws ( 3 – 8 ).
4.
5.
Remove the Monitor Cover L2 Assy (FRU 173) and R2 Assy (FRU 174).
6.
Remove the Tray2 Assy (FRU 175).
Loosen four screws ( 9 –
7.
8.
Remove the Escutcheon2 Assy (FRU 172).
9.
Disconnect one connector (
10. Disconnect one connector (
Monitor15 Cover Set includes the Monitor Cover L2 Assy (FRU 173), Monitor Cover R2 Assy
(FRU 174), Tray2 Assy (FRU 175), and the Escutcheon2 Assy (FRU 172).
12
).
13
).
14
) from the back side of the Escutcheon2 Assy.
Note
6–86
LOGIQ 500 SERVICE MANUAL