GE MEDICAL SYSTEMS
REV 14
6–2–19 Neck Space Plate (FRU No. 203)
Time Required
0.25 Hours
Tool Required
Screwdriver
Hexagonal Wrench
Procedure
Refer to ILLUSTRATION 6–32.
1.
Turn OFF the system.
2.
Remove the Monitor Assy (FRU No. 100) from the Neck Assy. Refer to 6–2–1 on page 6–66.
Remove the Neck Space Plate ( 1 ).
3.
Monitor
NECK SPACE PLATE DISASSEMBLY
ILLUSTRATION 6–32
6–95
LOGIQ 500 SERVICE MANUAL
1
Main body
RENEWAL PARTS
P9030TA