GE MEDICAL SYSTEMS
REV 14
1–3 EMC (Electromagnetic Compatibility)
1–3–1 EMC Performance
All types of electronic equipment may characteristically cause electromagnetic interference with other equipment, ei-
ther through air or connecting cables. The term EMC (Electromagnetic Compatibility) indicates capability of the equip-
ment, which curbs electromagnetic influence from other equipment and at the same time does not affect other equip-
ment with similar electromagnetic radiation from itself.
This product is designed to fully comply with the EN60601–1–2 (IEC601–1–2), in Medical electrical equipment EMC
regulations.
Manufacturer built consoles with version 2.00 software or later conform with EN60601–1–2.
Consoles upgraded to version 2.00 or later software (using optional upgrade kits) will not conform to
EN60601–1–2.
Proper installation following this service manual is required in order to achieve the full EMC performance of the prod-
uct.
The product must be installed as stipulated in 1–3–2, Notice upon Installation of Product.
In case of issues related to EMC, please follow procedures stated in 1–3–4, Countermeasures against EMC-related
Issues.
1–3–2 Notice upon Installation of Product
1) Use either power supply cords provided by GEMS or ones designated by GEMS. Products equipped with
power source plug should be plugged into the fixed power socket which has the protective grounding conduc-
tor.
Connect a three-pole plug to a three-pole socket without using a three-pole-to-two-pole converter.
2) Locate the equipment as far as possible from other electronic equipment.
3) Be sure to use either any cables provided by GEMS or ones designated by GEYMS. Wire these cables follow-
ing these installation procedures.
(Example) Wire power cables separately from signal cables.
4) Lay out the main equipment and other peripherals following the installation procedures described in Chapter2,
INSTALLATION.
Note