Time Keeper Ram (For Mvme167-002B) (Fru No. 616) - GE LOGIQ 500 Service Manual

Hide thumbs Also See for LOGIQ 500:
Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 14
6–2–78 Time Keeper RAM (for MVME167–002B) (FRU No. 616)
Time Required
1 Hour
Tool Required
Screwdriver
IC Remover
Procedure
Refer to ILLUSTRATION 6–108 on page 6–187.
1.
Turn OFF the system.
2.
Remove the Side Cover Left (FRU 301). Refer to 6–2–20 on page 6–97.
3.
Unscrew the four Nest Cover screws.
4.
Remove the Nest Cover.
5.
Remove the CPU–A Assy (FRU 610). Refer to 6–2–75 on page 6–180.
6.
Remove the Time Keeper RAM on the SBC–A Assy (MVME167–002B).
After replacing a time keeper RAM with new one, it is necessary to install the system ID number into
the time keeper RAM and to set up the factory preset data.
Note
6–186
LOGIQ 500 SERVICE MANUAL
P9030TA
RENEWAL PARTS