GE MEDICAL SYSTEMS
REV 14
1–2–2 Specifications
Type of protection against electric shock:
Degree of protection against electric shock: Type BF EQUIPMENT (*2) (Except ECG)
Ordinary Equipment
Continuous Operation
*1. Class I EQUIPMENT
EQUIPMENT in which protection against electric shock does not rely on BASIC INSULATION only, but which
includes an additional safety precaution in that means are provided for the connection of the EQUIPMENT to
the protective earth conductor in the fixed wiring of the installation in such a way that ACCESSIBLE METAL
PARTS cannot become LIVE in the event of a failure of the BASIC INSULATION.
*2. Type BF EQUIPMENT
TYPE B EQUIPMENT with an F–TYPE APPLIED PART
TYPE B EQUIPMENT: EQUIPMENT providing a particular degree of protection against electric shock, partic-
ularly regarding:
– allowable LEAKAGE CURRENT;
Patient leakage current
*3. Type CF EQUIPMENT
EQUIPMENT providing a particular degree of protection higher than that for TYPE OF BF EQUIPMENT
against electric shock particularly regarding allowable LEAKAGE CURRENT, and having an F–TYPE AP-
PLIED PART.
– allowable LEAKAGE CURRENT;
Patient leakage current
Class I EQUIPMENT (*1)
Type CF EQUIPMENT (*3) (ECG Only)
Normal mode
Single failure mode
Less than 100µA
Less than 500µA
Normal mode
Single failure mode
Less than 10µA
Less than 50µA
1–18
LOGIQ 500 SERVICE MANUAL
P9030TA