Mtz Probe Holder Installation; Transducer Connection - GE LOGIQ 500 Service Manual

Hide thumbs Also See for LOGIQ 500:
Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 7
2–2–6 MTZ Probe Holder Installation (Continued)
The standard probe holder is changed to new one from the LOGIQ 500 console with system software version 2.00.
Only one MTZ probe holder can be installed on the console with the new standard probe holder. Refer to
ILLUSTRATION 2–5.
2–2–7 Transducer Connection
1.
Connect a transducer to the upper transducer receptacle as follows:
a. Ensure that the transducer twist lock lever points towards the 3 o'clock position.
b. Insert the transducer connector on the receptacle guide pin until it touches the receptacle mating surface.
c. Twist the transducer twist lock lever to the 7 o'clock position to lock it in place. Twist the lever to the 3 o'clock
position to disconnect the transducer.
It is not necessary to turn OFF power to connect or disconnect a transducer.
2.
Connect the main power cable to a hospital grade power receptacle with the proper rated voltage checked during
preinstallation. Never use a three–to–two prong adapter; this defeats the safety ground.
MTZ PROBE HOLDER INSTALLATION (b)
ILLUSTRATION 2–5
Note
2–12
LOGIQ 500 SERVICE MANUAL
P9030TA
INSTALLATION