GE MEDICAL SYSTEMS
REV 14
6–2–96 Inrush Limiter 120V, Inrush Limiter 220V (FRU No. 721)
Time Required
30 Minutes
Tool Required
Screwdriver
Hexagonal Wrench
Procedure
Refer to ILLUSTRATION 6–126 on page 6–225.
1.
Turn OFF the system.
2.
Remove the Rear Cover Assy (FRU 303). Refer to 6–2–22 on page 6–99.
3.
Remove the Power Supply Unit. Refer to 6–5–2 on page 6–243.
Unscrew four screws ( 1 – 4 ). Refer to ILLUSTRATION 6–124 on page 6–221.
4.
5.
Remove the A7 SW Cont Assy (FRU No. 706). Refer to 6–2–88 on page 6–208.
Unscrew two nuts ( 5 and 6 ) and remove the two cables.
6.
7.
Remove the transparent plastic cover.
Unscrew two screws ( 7 and 8 ) and remove the two cables.
8.
Unscrew two screws ( 9 and 10 ).
9.
10. Pull two cables out by hand ( 11 and 12 ).
11. Remove the Inrush Limiter (FRU 721).
6–224
LOGIQ 500 SERVICE MANUAL
P9030TA
RENEWAL PARTS