Fdd (Fru No. 518), Fdd Holder Assy (Fru No. 519) - GE LOGIQ 500 Service Manual

Hide thumbs Also See for LOGIQ 500:
Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 14
6–2–69 FDD (FRU No. 518), FDD Holder Assy (FRU No. 519)
Time Required
15 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6–89.
1.
Turn OFF the system.
2.
Remove the Keyboard Panel Assy (FRU 400). Refer to 6–2–41 on page 6–121.
Disconnect two connectors ( 1 and 2 ).
3.
Unscrew four screws ( 3 – 6 ).
4.
5.
Remove the FDD Holder Assy.
Unscrew four screws ( 7 –
6.
7.
Remove the FDD (FRU 518) out of the FDD Holder Assy (FRU 519).
).
10
FDD AND FDD HOLDER ASSY DISASSEMBLY
ILLUSTRATION 6–89
6–161
LOGIQ 500 SERVICE MANUAL
P9030TA
RENEWAL PARTS