Running Auto-Teaching and Teach-Point Verification

Run auto-teaching

Auto-teaching is a process in which the system locates and records the positions of markers
(called teach points) printed on the empty deck, ensuring that, during a protocol, the micropipettor
is accurately aligned with the tubes or wells at each deck position.
NOT E
1 Before you start, make sure that the tube holder lids are closed and all tip boxes, strip tubes,
and plates have been removed from the deck.
2 From the Home screen, press Settings.
The
3 On the Settings screen, press Auto Teaching.
The
Figure 26 Auto Teach screen
4 Press Start.
A message box opens asking you to confirm that the tube holder lids are closed and all
plasticware has been removed from the deck.
5 Press OK to continue.
The system initiates auto-teaching.
6 When prompted on the touchscreen for a micropipettor tip, open the instrument door and add
a tip to the micropipettor at the indicated barrel position. Leaving the door open, press Next to
continue.
Magnis and MagnisDx NGS Prep System User Guide
At particular points during the auto-teaching process, the system prompts you
to add tips to the micropipettor and to place (and then later remove) a tip box
on the instrument deck.
Settings screen
opens.
Auto Teach screen
opens.