8.5.3 "Joystick" Parameter Settings

Menu tree
Input devices
> Hand control device settings
> Joystick
> Centre deadband
Input devices
> Hand control device settings
> Joystick
> Switch operation
Input devices
> Hand control device settings
> Joystick
> Tremor suppression
Table 8
"Joystick" parameter settings
Access
Min Max Spec. Unit Help text
Technical
10
50
20
Service
Technical
Off
On
Off
Service
Technical
0
100
0
Service
%
Centre deadband
Centre deadband defines the joystick
travel before the system recognises the
command. The value corresponds to a
circle around the joystick zero point.
No movements or commands are executed
unless the joystick is moved beyond this
circle.
Increasing this value is useful if the opera-
tor has a pronounced hand tremor.
O n /
Switch function
Off
The joystick can be used like a switch. If
the switch function is ON, the proportional
function is replaced by a switch function.
Moving the joystick from the centre by more
than 50% generates a 100% command
in the desired direction.
%
Tremor suppression: Suppresses a tremor
on the joystick.
0 % -> Tremor suppression deactivated
100 % -> Maximum tremor suppression
If a brief command is used
INFORMATION
for system operation (e.g. if "3 direction
profile" is selected), tremor suppression
should not exceed 90%.
If tremor suppression exceeds 90% in
such a case, the brief command is ignored.