Brake Settings; Performing A Function Check Of The Brake Settings - Otto Bock Xeno Service Manual

Otto bock healthcare xeno power wheelchair
Hide thumbs Also See for Xeno:
Table of Contents

Advertisement

Figure 41 Adjusting the Bowden cable
1 Adjustment screws on the brake release housing
2 Threaded rod for readjusting the drive motor

6.6.3 Brake Settings

CAUTION
Risk of accidents and injury due to lack of brake functionality. Incorrect brake settings can lead to
a loss of braking power. Repairs and settings must only be made by authorised personnel. Replace
damaged Bowden cables with new ones immediately.
Correct brake settings are essential for driving. This is why Otto Bock recommends inspecting
the brake every three months.
To avoid excessive friction, the radius of the Bowden cable must not be less than 10 cm.
Damaged Bowden cables must be replaced immediately, for example if individual wires are
protruding or there is a kink in the black sheath.

6.6.4 Performing a function check of the brake settings

During the functional test of the brake settings, please observe the following points and complete
the tasks described below:
Steps:
3. Lock the release lever in position 1.
4. While driving straight forward at approximately 6 km/h, release the joystick in its front-most
position.
5. If the wheelchair maintains its current direction during the delay, repeat the test on a slope of
10% unless this is limited by information on the nameplate.
6. If the power wheelchair leaves the current direction, adjust the Bowden cable for the opposite
side. The adjustment process is described in Section 6.6.2 under "Adjusting an over-tightened
Bowden cable".
7. If the power wheelchair tracks properly in both cases, conduct the following test on an even
surface:
a. Pull the release bolt on the brake housing. The lever automatically moves forward into posi-
2
1
43