Otto Bock Xeno Service Manual page 91

Otto bock healthcare xeno power wheelchair
Hide thumbs Also See for Xeno:
Table of Contents

Advertisement

Main
Menu
Parameters
System
General drive
settings
Rescue drive
Wheelchair data
Reverse beep
Timing
Motor & wheels
Registered office
Submenu
Settings
Max speed
The maximum speed parameter forms another layer
above the speed values that can be programmed under
input devices.
Example: Max. speed = 60%.
Forward speed = 80%. In driving mode 1: 0.8 x 60% =
48% of the maximum driving speed.
Maximum RPM
Limits the maximum turning speed.
Quick stop fwd
Factor that determines the deceleration, e.g. if the joystick
factor
is released while driving.
Quick stop
rev factor
Emergency stop
Parameter that determines the deceleration in case of
problems in the control unit, e.g. the wheelchair is turned
off while driving.
Forward tracking
Compensates for speed differences between the right-hand
and left-hand motor. Directional stability can be improved.
Reverse tracking
V-steering limit
Determines speed changes in curves.
Mode button
Defines whether long and short mode signals have dif-
navigation
ferent meanings.
Profile increase
stop
All drive parameters if the wheelchair is in an emergency
situation (creep speed).
Parameters that relate to the wheelchair base frame,
e.g. wheelbase. They should not be changed.
A warning signal can be emitted during reverse operation.
It can be activated or deactivated in the individual profiles.
Various times can be programmed.
Automatic shutoff The wheelchair turns off after the time set here. It does
not shut down if the setting is 0.
Latch timeout
After the time specified here, the wheelchair switches
from cruise control operation to normal operation.
Latch timeout
Specifies whether the cruise control automatically stops
deactivated
after the latch timeout or not.
Drive timeout
Time between the last driving command and automati-
cally switching to the prior mode. No switching if the
parameter is set to 0.
Aux timeout
Time between the last input command and automatically
switching to the prior operating mode. No switching if
the parameter is set to 0.
Menu timeout
Defines the command time for menu operation.
IR function
Time before the IR menu jumps up one level. No jump if
timeout
the parameter is set to 0.
Parameters that adapt the drive motors to the controller.
The direction of rotation can be changed.
Actuators 1 to 5
Parameters to adjust the seat actuators (speed, current).
The seat module offers a separate 12 volt output. The
voltage can be turned on or off under the "Miscellane-
ous" menu item.