94
Access
Min Max Spec. Unit Help text
Technical
5
100
20
Service
Technical
5
100
20
Service
Technical
5
100
20
Service
Technical
5
100
20
Service
%
Maximum forward acceleration
Specifies the forward acceleration of
the wheelchair when the position of the
potentiometer is set to maximum.
A high value represents a short accelera-
tion time and a fast start. High acceleration
values result in abrupt speed changes
and should only be used in special cases.
If the position of the potentiometer is
between minimum and maximum, the
forward acceleration is adjusted linearly.
Ensure the user can operate the wheelchair
safely at maximum acceleration.
%
Maximum forward deceleration
Specifies the forward deceleration of
the wheelchair when the position of the
potentiometer is set to maximum.
A high value represents a short decelera-
tion time and a rapid stop. High decelera-
tion values result in abrupt speed changes
and should only be used in special cases.
If the position of the potentiometer is
between minimum and maximum, the
deceleration is adjusted linearly.
Ensure the user can operate the wheelchair
safely at maximum deceleration.
%
Maximum reverse acceleration
Specifies the reverse acceleration of
the wheelchair when the position of the
potentiometer is set to maximum.
A high value represents a short accelera-
tion time and a fast start. High acceleration
values result in abrupt speed changes
and should only be used in special cases.
If the position of the potentiometer is
between minimum and maximum, the
reverse acceleration is adjusted linearly.
Ensure the user can operate the wheelchair
safely at maximum acceleration.
%
Maximum reverse deceleration
Specifies the reverse deceleration of
the wheelchair when the position of the
potentiometer is set to maximum.
A high value represents a short decelera-
tion time and a rapid stop. High decelera-
tion values result in abrupt speed changes
and should only be used in special cases.
If the position of the potentiometer is
between minimum and maximum, the
reverse delay is scaled linearly.
Ensure the user can operate the wheelchair
safely at maximum deceleration.