6.16.5 Other Optional Add-on Components

INFORMATION
These and other optional add-on components are included in the order form and in the accessories
catalogue.
INFORMATION
A rear cover can only be mounted in combination with the optional lighting.
❒ Luggage carrier
❒ Puncture resistant tyres: solid rubber tyres
❒ Tray: attachable tray top
❒ Pocket for mobile phone

7 Diagnostics

NOTICE
Risk of damage due to unauthorised service. Improper or poorly executed repairs may result in the
unsafe operation of the power wheelchair. The error diagnosis may therefore only be carried out by
authorised dealers who have established knowledge of electronic controllers from Otto Bock.
Otto Bock will not assume any liability for damages that are due to improperly or poorly executed repairs.
INFORMATION
Experience has shown that problems with the wheelchair electrical system are frequently caused by
errors and defects in the plug connections and cabling. They should be inspected first for this reason.
INFORMATION
Should you encounter problems while troubleshooting or if you do not manage to completely eliminate
a problem by following the measures described here, please contact Otto Bock.
Malfunctions are either shown on the control panel display/LCD monitor (see Section 7.2) or dis-
played using the handheld programming device. The handheld programming device (see Section
8.4) can capture errors more precisely using error codes. The separate enAble50 (647G490=D/
GB) instructions for use contain a table with error codes, causes and possible steps for resolution.
7.1 Diagnostic Steps
For an efficient and successful error diagnosis, proceed as follows:
❒ Verify whether the error was generated for a concrete reason or if the error was displayed
sporadically.
❒ Check the possible causes in the error display table (see Section 7.2) or the error codes
(enAble50 (647G490=D/GB instructions for use). Perform the recommended checks and
corrective actions.