Otto Bock Xeno Service Manual page 96

Otto bock healthcare xeno power wheelchair
Hide thumbs Also See for Xeno:
Table of Contents

Advertisement

Menu tree
Input devices
> Hand control device settings
> Drive
> Profile 1P1
> Latch Reverse LatchRev
Input devices
> Hand control device settings
> Drive
> Profile 1P1
> Power
Table 7
"Drive" parameter settings
96
Access
Min Max Spec. Unit Help text
Technical
0
2
0
Service
Technical
20
100
100
Service
Constant speed forward / reverse
If „Constant speed", cruise, is selected, the
wheelchair accelerates to the desired speed.
If the command is reduced, the wheelchair
maintains the speed. Direction changes are
always possible. The wheelchair decelerates
if an opposing command of 50% or less
is issued. A stop is achieved via a mode
button command or an opposing command
of more than 50%.
If „Constant speed", stepped, is selected,
the wheelchair accelerates by one incre-
ment with each short command (33%, 66%
and 100%) up to the maximum speed.
Direction changes are always possible.
The wheelchair decelerates by one incre-
ment (e.g. from 66% to 33%) when a brief
opposing command of 50% is issued. A
stop is achieved via a mode switch or with
a permanent opposing command.
Risk of injury due to incor-
WARNING
rectly installed control buttons.
Never select "Constant speed" mode
(cruise or stepped) unless the mode button
is installed so that the user can operate
it quickly at any time.
%
Performance
The power parameter may be less than
100% to limit the maximum torque.
Limiting the torque reduces the power of
the wheelchair. This can prevent dangerous
collisions or crossing obstacles.
Reducing torque usually only makes sense
for indoor operation.
After reducing the torque, ensure the user
is able to cross any obstacles that need
to be crossed.