Actuator Outputs; Seat Module - Otto Bock Xeno Service Manual

Otto bock healthcare xeno power wheelchair
Hide thumbs Also See for Xeno:
Table of Contents

Advertisement

8.2.3 Actuator outputs

The controller has two independent actuator outputs.
Technical data for the first actuator output:
Voltage
Current for 20 seconds
Steady current
Technische Daten des zweiten Ausgangs:
Voltage
Current for 20 seconds
Steady current
The actuator outputs are short-circuit protected. However, short-circuiting the actuator output
against the negative pole of the battery while the actuator is in operation may damage the output.
The actuators have high impedance when they are not in operation.

8.3 Seat module

The controller is attached to the wheelchair with two screws and can be aligned in any position
(see Section 6.12.5).
1
3
Figure 91 AAM
1 IN-A, IN-B, IN-C, IN-D, IN-E connections
2 RSW
3 A1 - A5 connections (actuators A1 - A5)
4 BUS connection
5 BUS connection
6 BATT bridge/external supply connection
7 12 V DC (connection must be activated by parameter -> On/Off!)
84
2
4
5 6 7
0...100 % V
Batt
20 A
10 A
0 ... 100 % V
Batt
15 A
7 A

Advertisement