8.6.3 "Joystick" parameter settings

Menu tree
Input Devices
> Handcontrol Settings
> Joystick
> Center deadband
Input Devices
> Handcontrol Settings
> Joystick
> Switch operation
Input Devices
> Handcontrol Settings
> Joystick
> Tremor suppression
Skippi Plus
enAble50 wheelchair control – installation and programming
Access Min. Max.
De-
fault
Service
10
50
Service
Off
On
Service
0
100
Unit
20
%
Center deadband
Center deadband defines the joystick
travel before the system recognises
the command. The value corresponds
to a circle around the joystick zero
point.
No movements or commands are
executed unless the joystick is moved
beyond this circle.
Increasing this value is useful if the
operator has a pronounced hand
tremor.
Off
On/
Switch function
Off
The joystick can be used like a switch.
If the switch function is ON, the
proportional function is replaced by a
switch function.
Moving the joystick from the centre by
more than 50% generates a 100%
command in the desired direction.
0
%
Suppresses a tremor on the joystick.
0% -> Tremor suppression deactiva-
ted
100% -> Maximum tremor suppres-
sion
INFORMATION
used for system operation (e.g. if "3
direction profile" is selected), tremor
suppression should not exceed 90%.
If tremor suppression exceeds 90%
in such a case, the short command is
ignored.
Help text
If a short command is
Ottobock | 65