Otto Bock Xeno Service Manual page 22

Otto bock healthcare xeno power wheelchair
Hide thumbs Also See for Xeno:
Table of Contents

Advertisement

Figure 9
Charging receptacle
1 Receptacle for connecting the charger
2 Cover
Figure 10 Otto Bock battery charger
The battery charger features a programmed recharging phase. Once a discharged battery is
completely charged (after approximately 8 hours), the battery charger can remain connected
with no risk of overcharging or damaging the battery.
The battery charger indicates the following states:
Anzeige
Green LED flashes
Yellow LED is lit
Green LED is illuminated
Red LED is illuminated
Table 2
Battery charging statuses during the charging process
22
Funktion
1 x
Low-maintenance battery
2 x
Maintenance-free battery
Steady
Battery is charged to 80 %
Battery is charging
Battery is fully charged
Battery is defective or missing, or charging time has been
exceeded
2
1

Advertisement

Table of Contents
loading

Table of Contents