6.9 Registered office
CAUTION
Risk of burns in the proximity of fire. The back rest, seat bottom and supports of the XENO power
wheelchair are flame retardant but nevertheless flammable. Therefore utmost caution is required near
any sources of open flame or sparks, especially lit cigarettes.
NOTICE
Risk of damage to upholstery. During any work on the seat, the upholstery must be adequately pro-
tected against mechanical and chemical damage.

6.9.1 Removing the Seat Cushion and Back Padding

The seat cushion and back padding are attached to the seat and back plates with two hook-
and-loop closures and can simply be pulled away.

6.9.2 Changing the backrest height setting

NOTICE
Risk of cable damage. Ensure the cables attached to the frame with cable ties are not damaged.
Preparation:
❒ Remove the backrest cushion from the back plate (see Section 6.9.1).
Tools:
❒ Ring wrench or open-end wrench, size 10 mm
Steps:
1. Loosen the four Allen head screws on the back side of the back plate.
2. Slide the back plate along the slotted holes to set the height.
Properly reassemble all components upon completion of the work.