Table 1
Information displays

6 Service and Maintenance

6.1 Inspecting the General Condition
❒ Check all safety-related components for corrosion, repair if required and reapply corrosion
protection.
❒ Check weld seams.
❒ Check tightness of screw connections; replace thread-locking compound if required; observe
defined torque specifications; replace defective screws (e. g. in case of corrosion).
❒ Check cables for ruptures, signs of wear and proper attachment; replace defective components.
❒ Check cable connections and plug connections.

6.2 Fuse replacement

The 80 A fuse is located in a fuse holder at the rear end of the battery case.
Preparation:
❒ OWithout lighting: Pull the battery case out to the rear and remove the cover.
❒ With lighting: Pull the battery case out to the rear and remove the cover plate (see Section
6.3.2).
❒ Turn the control unit off.
Steps:
1. Open the cap of the fuse holder.
2. Pull out the fuse and replace if required.
3. When inserting the fuse, make sure that it is pressed into the centre of the spring contacts
and that it is not seated at an angle.
4. Close the cover. You should feel the cover lock in place.
Properly reassemble all components upon completion of the work.