8719ET/20ET/22ET
8719ES/20ES/22ES
RF Network Analyzers
Figure 3-13 Setup for the R Channel (ET/ES Models)

Raw Offset for the R Channel (ET/ES Models)

1. Press the following:
Preset
FACTORY PRESET
SERVICE MENU
System
Stop
10
G/n
INPUT PORTS
Meas
Marker Search
SEARCH: MAX
Note the maximum frequency and write it down.
2. Press
Sweep Setup
Press
G/n
Marker Search
Note the maximum marker reading (MMR) and write it down.
3. Enter the power sensor calibration factor for the maximum frequency point into the
power meter.
NOTE
Refer to the power sensor manual for instructions on how to enter a
calibration factor into the power meter. If the exact frequency is not listed on
the power sensor, use a value that is between the two nearest points.
Service Guide
Preset
SERVICE MODES
R
, and enter the maximum frequency point from step 1.
CW FREQ
SEARCH: MAX
Adjustments and Correction Constants
Raw Offset Correction Constants
MORE
SAMPLER COR OFF
.
3- 49