:MARKer:MODE

Command Syntax
:MARKer:MODE <mode>
<mode> ::= {OFF | MEASurement | MANual | WAVeform}
The :MARKer:MODE command sets the cursors mode:
• OFF — removes the cursor information from the display.
• MANual — enables manual placement of the X and Y cursors.
• MEASurement — cursors track the most recent measurement.
• WAVeform — the Y1 cursor tracks the voltage value at the X1 cursor of
Query Syntax
:MARKer:MODE?
The :MARKer:MODE? query returns the current cursors mode.
Return Format
<mode><NL>
<mode> ::= {OFF | MEAS | MAN | WAV}
See Also
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
(see
page
790)
If the front- panel cursors are off, or are set to the front- panel Hex or
Binary mode, setting :MARKer:MODE MANual will put the cursors in
the front- panel Normal mode.
Setting the mode to MEASurement sets the marker sources
(:MARKer:X1Y1source and :MARKer:X2Y2source) to the measurement
source (:MEASure:SOURce). Setting the measurement source remotely
always sets the marker sources.
the waveform specified by the X1Y1source, and the Y2 cursor does the
same for the X2 cursor and its X2Y2source.
"Introduction to :MARKer Commands"
":MARKer:X1Y1source"
":MARKer:X2Y2source"
":MEASure:SOURce"
":MARKer:X1Position"
":MARKer:X2Position"
":MARKer:Y1Position"
":MARKer:Y2Position"
on page 318
on page 321
on page 323
on page 361
on page 320
on page 322
on page 327
on page 328
19
:MARKer Commands
319

Advertisement

Table of Contents
loading

Table of Contents