:WGEN Commands
Waveform
Type
PULSe
DC
NOISe
For all waveform types, the output amplitude, into 50
from 10 mVpp to 2.5 Vpp (or from 20 mVpp to 5 Vpp into and
open- circuit load).
Query Syntax
:WGEN:FUNCtion?
The :WGEN:FUNCtion? query returns the currently selected signal type.
Return Format
<signal><NL>
<signal> ::= {SIN | SQU | RAMP | PULS | NOIS | DC}
See Also
668
Characteristics
Use these commands to set the pulse signal parameters:
":WGEN:FREQuency"
":WGEN:PERiod"
":WGEN:VOLTage"
":WGEN:VOLTage:OFFSet"
":WGEN:VOLTage:HIGH"
":WGEN:VOLTage:LOW"
":WGEN:FUNCtion:PULSe:WIDTh"
The frequency can be adjusted from 100 mHz to 10 MHz.
The pulse width can be adjusted from 20 ns to the period minus 20 ns.
Use this command to set the DC level:
":WGEN:VOLTage:OFFSet"
Use these commands to set the noise signal parameters:
":WGEN:VOLTage"
":WGEN:VOLTage:OFFSet"
":WGEN:VOLTage:HIGH"
":WGEN:VOLTage:LOW"
"Introduction to :WGEN Commands"
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
on page 666
on page 686
on page 688
on page 691
on page 689
on page 690
on page 669
on page 691
on page 688
on page 691
on page 689
on page 690
on page 665
, can be adjusted