Acquire:complete - Agilent Technologies InfiniiVision 2000 X-Series Programmer's Manual

Oscilloscopes
Hide thumbs Also See for InfiniiVision 2000 X-Series:
Table of Contents

Advertisement

:ACQuire:COMPlete

Command Syntax
:ACQuire:COMPlete <complete>
<complete> ::= 100; an integer in NR1 format
The :ACQuire:COMPlete command affects the operation of the :DIGitize
command. It specifies the minimum completion criteria for an acquisition.
The parameter determines the percentage of the time buckets that must be
"full" before an acquisition is considered complete. If :ACQuire:TYPE is
NORMal, it needs only one sample per time bucket for that time bucket to
be considered full.
The only legal value for the :COMPlete command is 100. All time buckets
must contain data for the acquisition to be considered complete.
Query Syntax
:ACQuire:COMPlete?
The :ACQuire:COMPlete? query returns the completion criteria (100) for
the currently selected mode.
Return Format
<completion_criteria><NL>
<completion_criteria> ::= 100; an integer in NR1 format
See Also
Example Code
See complete example programs at:
starting on page 799
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
(see
page
790)
"Introduction to :ACQuire Commands"
":ACQuire:TYPE"
on page 195
":DIGitize"
on page 159
":WAVeform:POINts"
on page 640
' AQUIRE_COMPLETE - Specifies the minimum completion criteria for
' an acquisition.
The parameter determines the percentage of time
' buckets needed to be "full" before an acquisition is considered
' to be complete.
myScope.WriteString ":ACQuire:COMPlete 100"
:ACQuire Commands
on page 183
Chapter
38, "Programming Examples,"
7
185