:CHANnel<n>:PROBe
:CHANnel<n>:PROBe <attenuation>
<attenuation> ::= probe attenuation ratio in NR3 format
<n> ::= 1 to (# analog channels) in NR1 format
The obsolete attenuation values X1, X10, X20, X100 are also supported.
The :CHANnel<n>:PROBe command specifies the probe attenuation factor
for the selected channel. The probe attenuation factor may be 0.1 to
10000. This command does not change the actual input sensitivity of the
oscilloscope. It changes the reference constants for scaling the display
factors, for making automatic measurements, and for setting trigger levels.
Query Syntax
:CHANnel<n>:PROBe?
The :CHANnel<n>:PROBe? query returns the current probe attenuation
factor for the selected channel.
Return Format
<attenuation><NL>
<attenuation> ::= probe attenuation ratio in NR3 format
See Also
.
See complete example programs at:
starting on page 799
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
(see
page
790)
":CHANnel<n>:OFFSet"
' CHANNEL_PROBE - Sets the probe attenuation factor for the selected
' channel.
The probe attenuation factor may be set from 0.1 to 10000
myScope.WriteString ":CHANnel1:PROBe 10"
on page 226
Chapter
38, "Programming Examples,"
' Set Probe to 10:1.
10