24
:SAVE Commands

:SAVE:WMEMory:SOURce

Command Syntax
:SAVE:WMEMory:SOURce <source>
<source> ::= {CHANnel<n> | FUNCtion | MATH | WMEMory<r>}
<n> ::= 1 to (# analog channels) in NR1 format
<r> ::= {1 | 2}
The :SAVE:WMEMory:SOURce command selects the source to be saved as a
reference waveform file.
Only ADD or SUBtract math operations can be saved as reference waveforms.
N O T E
MATH is an alias for FUNCtion. The query will return FUNC if the source is FUNCtion or
N O T E
MATH.
Query Syntax
:SAVE:WMEMory:SOURce?
The :SAVE:WMEMory:SOURce? query returns the source to be saved as a
reference waveform file.
Return Format
<source><NL>
<source> ::= {CHAN<n> | FUNC | WMEM<r> | NONE}