Marker:x2Y2Source - Agilent Technologies InfiniiVision 2000 X-Series Programmer's Manual

Oscilloscopes
Hide thumbs Also See for InfiniiVision 2000 X-Series:
Table of Contents

Advertisement

:MARKer:X2Y2source

Command Syntax
:MARKer:X2Y2source <source>
<source> ::= {CHANnel<n> | FUNCtion | MATH | WMEMory<r>}
<n> ::= 1 to (# analog channels) in NR1 format
<r> ::= {1 | 2}
The :MARKer:X2Y2source command sets the source for the cursors. The
channel you specify must be enabled for cursors to be displayed. If the
channel or function is not on, an error message is issued.
If the marker mode is not currently WAVeform (see
page 319):
• Sending a :MARKer:X2Y2source command will put the cursors in the
• Setting the source for one pair of markers (for example, X2Y2) sets the
If the marker mode is currently WAVeform, the X2Y2 source can be set
separate from the X1Y1 source.
If :MARKer:MODE is set to OFF or MANual, setting :MEASure:SOURce to
CHANnel<n>, FUNCtion, MATH, or WMEMory<r> will also set
:MARKer:X1Y1source and :MARKer:X2Y2source to this value.
MATH is an alias for FUNCtion. The query will return FUNC if the source is FUNCtion or
N O T E
MATH.
Query Syntax
:MARKer:X2Y2source?
The :MARKer:X2Y2source? query returns the current source for the
cursors. If all channels are off or if :MARKer:MODE is set to OFF, the
query returns NONE.
Return Format
<source><NL>
<source> ::= {CHAN<n> | FUNC | WMEM<r> | NONE}
See Also
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
(see
page
790)
MANual mode.
source for the other (for example, X1Y1).
"Introduction to :MARKer Commands"
":MARKer:MODE"
on page 319
":MARKer:X1Y1source"
":MEASure:SOURce"
on page 318
on page 321
on page 361
:MARKer Commands
":MARKer:MODE"
on
323
19

Advertisement

Table of Contents
loading

Table of Contents