Wmemory<R>:Yoffset - Agilent Technologies InfiniiVision 2000 X-Series Programmer's Manual

Oscilloscopes
Hide thumbs Also See for InfiniiVision 2000 X-Series:
Table of Contents

Advertisement

32
:WMEMory<r> Commands
:WMEMory<r>:YOFFset
Command Syntax
:WMEMory<r>:YOFFset <offset> [<suffix>]
<r> ::= 1-2 in NR1 format
<offset> ::= vertical offset value in NR3 format
<suffix> ::= {V | mV}
The :WMEMory<r>:YOFFset command sets the value that is represented at
center screen for the selected reference waveform.
The range of legal values varies with the value set by the
:WMEMory<r>:YRANge or :WMEMory<r>:YSCale commands. If you set the
offset to a value outside of the legal range, the offset value is
automatically set to the nearest legal value. Legal values are affected by
the probe attenuation setting.
Query Syntax
:WMEMory<r>:YOFFset?
The :WMEMory<r>:YOFFset? query returns the current offset value for the
selected reference waveform.
Return Format
<offset><NL>
<offset> ::= vertical offset value in NR3 format
See Also
700
(see
page
790)
Chapter
32, ":WMEMory<r> Commands," starting on page 693
":WMEMory<r>:DISPlay"
":WMEMory<r>:YRANge"
":WMEMory<r>:YSCale"
":WMEMory<r>:SKEW"
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
on page 696
on page 701
on page 702
on page 699