Setting Offset Values
Set an offset value when the reference value for zero reset is a value other than 0.
Moving to FUN and SPCL
1.
Set the mode switch to FUN.
2.
Use the LEFT and RIGHT Keys to display
[SPCL] on the main display.
Moving to ZRDSP
3.
Press the UP or DOWN Key.
The sub-display will flash.
4.
Use the UP and DOWN Keys to display
[ETC] or [ALL].
5.
Press the ENT Key.
6.
Use the LEFT and RIGHT Keys to display
[ZRDSP] on the main display.
Setting Offset Values
7.
Press the UP or DOWN Key.
The leftmost digit of the sub-display will flash.
8.
Set the desired offset value.
9.
Press the ENT Key to confirm the setting.
The setting will be registered.
Section 6