Section 5

DETAILED SETTINGS

Setting Number of Samples to Average

The average of the measured values obtained based on the preset number of samples can be
output. This setting can be used when you want to ignore rapid changes in measured values.
If this setting is made, response time for judgement and linear outputs will drop.
Selection for No. of
samples to average
1
2
4
8
16
32
64
128
256
512
1024
2048
4096
Moving to FUN mode and AVE
1.
Set the mode switch to FUN.
2.
Use the LEFT and RIGHT Keys to display
[AVE] on the main display.
Selecting Number of Samples to Average
3.
Press the UP or DOWN Key.
The sub-display will flash.
4.
Use the UP and DOWN Keys to select the
desired number of samples to average.
5.
Press the ENT Key to confirm the setting.
The setting will be registered.
ZX-L-N
66
User's Manual
In the case of Reflective Type Sensor Head, the
Response time
number of samples to average has been set to
256 prior to shipment from the factory.
0.3 ms
In the case of Through-beam Type Sensor Head,
0.5 ms
0.8 ms
the number of samples to average has been set to
1.5 ms
32 prior to shipment from factory.
2.5 ms
5 ms
10 ms
20 ms
40 ms
75 ms
150 ms
300 ms
600 ms
In general, when the number of samples to average is
multiplied by "n", the resolution will increase by
SUB
SUB
SUB
n
.
LD ON
ZERO
ENABLE
(mm)