Omron L User Manual page 116

Smart sensors laser type
Table of Contents

Advertisement

Correcting Linear Output Values
Discrepancies may occur between the linear output current (or voltage) values set on
the Amplifier Unit and the actual current (or voltage) values measured due to the
conditions for the connected external device or other factors. The linear output
correction function can be used to correct this discrepancy.
The output values are corrected by entering the correction value for the current or
voltage values for any two points.
Set the monitor focus function and select either current or voltage output beforehand.
p.95
This section uses a current output as an example. Change the values in this example
for voltage output as necessary.
1.
Connect the linear output to an external ammeter.
Moving to FUN Mode and SPCL
2.
Turn ON the power supply and move the
mode switch to FUN.
3.
Use the LEFT and RIGHT Keys to display
[SPCL] on the main display.
Moving to LEFT-ADJ
4.
Press the UP or DOWN Key.
The sub-display will flash.
5.
Use the UP and DOWN Keys to display [SET]
or [ALL].
Section 5
DETAILED SETTINGS
LD ON
ZERO
ENABLE
(mm)
SUB
SUB
ZX-L-N
User's Manual
99

Advertisement

Table of Contents
loading

This manual is also suitable for:

ZxZx-l-n series

Table of Contents