Top Cover (Lower Section) Replacement Procedures

8-2-6-1
Tools
Use the appropriate Phillips-type screw drivers, and a wire cutter as indicated in the replacement
procedures for the lower section of the top cover.
8-2-6-2
Preparations
Shut down the Vivid 3N Pro/Expert ultrasound unit, as described in
8-2-6-3
Top Cover Lower Section Removal Procedure
1) Remove the side covers, as described in
4.
2) Remove the rear cover, as described in
3) Raise the control console to its maximum height.
4) Remove the two screws located in the brackets underneath the lower section of the top cover, one
on either side of the unit, as shown below:
5) Remove the connectors panels cover, as described the
Replacement Procedures" on page
6) Unplug all the peripheral cables connected to the rear connector panels.
NOTE:
Pay attention to the location of each cable, as each one must be reconnected to its
original location during the installation process.