LCD Monitor Adjustments
6-5-1
Overview
This procedure describes how to adjust the LCD monitor for use with Vivid 7, software version v6.x and
newer. It also describes how to set up the Vivid 7 to the factory default for LCD use.
The LCD adjustment procedure contains two main tasks:
1.) Adjust LCD monitor
2.) Select gamma curve
6-5-2
Cautions and Warnings
N/A
6-5-3
Access to Adjustments
The adjustments of the LCD monitor is done via controls (buttons) on front of the LCD monitor.
On the LCD monitor:
Press Menu to activate the OnsScreenDisplay (OSD) menu.
•
Navigate by using the up/down arrow buttons.
•
Select by using the Return button.
•
Back up in the menu tree using the Menu key.
•
The AUTO and PIP keys should not be used.