4.
Attach the Stabilizing Membrane.
5.
Plug the power cord into the Power Entry Module/Power Cord Socket and switch the power on.
6.
Power the system on by pressing the Power Button on the front of the system.
7.
Perform system functional checks listed in the Functional Checks section.
Transducer Functional Check
B-Mode Functional Checks
3D Volume Functional Checks
Invenia ABUS 2.0 – System Setup and Basic Service Manual
4700-0043-00 Rev. 4
Ensure that the transducer:
•
Upon Power Up moves across the Scan Head Assembly.
•
That the lights on the Scan Head Assembly light up.
•
That the transducer images as a scan is being performed.
•
That the transducer generates a 3D Volume.
Confirm that the system:
•
Performs a B-Mode scan.
•
Places B-Mode imaging parameters on the Touch Screen
Monitor.
•
Images evenly across the entire Touch Screen.
Confirm that the 3D Volume:
•
Gets generated after each Scan View performed.
Replacing the Transducer
8-11