Performing analog adjustment
Please refer to 4.3.4.10. Analog output adjustment.
1. In the Measuring Mode press SET, select main menu
with or and confirm selection with SET.
Each channel is adjusted at three points in the analog
range (at 10 %; 50 %; 90°% of the analog scale).
2. Select
Analog Adj. Ch. 1
3. Use or to select
4. Press SET. Read off multimeter display (e.g. 5601 mA) and
enter this value in the user menu. Do this by scrolling one digit
to the right using and increasing the value of digit by 1 using
. Confirm with
5. Use or to select
6. Press SET. Read off multimeter display (e.g. 12,001 mA) and
enter this value in the user menu. Do this by scrolling one digit
to the right using and increasing the value of digit by 1 using
. Confirm with
7. Use or to select
8. Press SET. Read off multimeter display (e.g. 18,401 mA) and
enter this value in the user menu. Do this by scrolling one digit
to the right using and increasing the value of digit by 1 using
. Confirm with
9. With or continue to
to 8).
10.Return to main menu
11.Continue to main menu
Measuring Mode with ESC.
Performing the pressure adjustment
Please refer to 4.3.4.11. n-point adjustment (pressure)
1. In the Measuring Mode press SET, select main menu
with or and confirm selection with SET.
The pressure adjustment can be performed at 3 to 6
adjustment points.
2. Select
Adjust Pressure
3. Use or to select
4. Press SET. Read off pressure sensor display (e.g. 30.1 Pa) and
enter this value in the user menu. Do this by scrolling one digit
to the right using and increasing the value of digit by 1 using
. Confirm with
5. Repeat steps 1 to 4 for adjustment points 2 to 6.
6. R eturn to main menu
with or and confirm with SET.
Adj. Point
1.
SET
or abort entry with ESC.
Adj. Point
2.
SET
or abort entry with ESC.
Adj. Point
3.
SET
or abort entry with ESC.
Analog Adj. Ch.
Adjust
with ESC.
Reset
with or or return to
with or and confirm with SET.
Adj. Point
1.
SET
or abort entry with ESC.
Adjust
with ESC.
4 Transmitter
Adjust
2or
3
(repeat steps 3
Adjust
79