TESTO 6381 Instruction Manual

TESTO 6381 Instruction Manual

Differential pressure transmitter
Hide thumbs Also See for 6381:
Table of Contents

Advertisement

Quick Links

testo 6381 · differential pressure transmitter
P2A software · Parameterizing, adjusting and analyzing
software
Instruction manual

Advertisement

Table of Contents
loading

Summary of Contents for TESTO 6381

  • Page 1 6381 · differential pressure transmitter P2A software · Parameterizing, adjusting and analyzing software Instruction manual...
  • Page 3: Safety And The Environment

    Pos : 8 /TD /Sic herheit und Umw elt/ Sic her heit gew ährl eist en/MU F 63xx /F achpersonal @ 3\m od_1234794940409_79.docx @ 26337 @ @ 1 Any additional work must only be carried out by authorized personnel. Otherwise testo will not accept any responsibility for the proper functioning of the instrument after repair and for the validity of certifications.
  • Page 4: About This Document

    2 About this document Pos : 11 /TD/Ü berschriften/MUF /1 Zu di esem Dokum ent @ 3\mod_1234793991331_79.docx @ 26242 @ 1 @ 1 About this document Pos : 12 /TD/Sic her heit und Umw elt/Z u dies em D ok ument /Verwendung/Verwendung (Standard) @ 0\m od_1173775068554_79. docx @ 337 @ 5 @ 1 >...
  • Page 5: Table Of Contents

    3 Contents Pos : 17 /TD/Ü berschriften/MUF /3 Inhalt @ 3\m od_1234794019831_79. docx @ 26261 @ 1 @ 1 Contents Safety and the environment ............. 3 About this document ..............4 Contents .................. 5 Transmitter ................8 4.1. Specifications ................ 8 4.1.1.
  • Page 6 Password protection ....................34 4.4.4. Structure of user menu ....................35 4.4.5. Overview of the testo 6381 user menu ..............36 4.4.6. The individual main menus ..................39 4.4.6.1. Editing main menu of channel 1 ..............39 4.4.6.2. Editing Main Menu Alarm ................40 4.4.6.3.
  • Page 7 Tips and assistance ..............82 6.1. Questions and answers ............82 6.2. Accessories and spare parts ..........82 6.2.1. Ordering options for 6381 transmitter (0555 6381)........... 83 Pos : 18 /TD/--- Seit enwec hs el --- @ 0\m od_1173774430601_0.docx @ 283 @ @ 1...
  • Page 8: Transmitter

    Pos : 22 /TD/Leistungs bes chr eibung/Liefer umf ang/MUF 63xx /MUF 63x x ohne Ethernet @ 3\m od_1234444980038_79.docx @ 25098 @ 3 @ 1 4.1.2. Scope of delivery The scope of delivery of the testo 6381 transmitter includes the following: •...
  • Page 9: Technical Data

    4 Transmitter 4.1.4. Technical data Parameters • Differential pressure Accuracy The specifications are only valid if the positive pressure is applied at the positive pressure connection. • 0.5 % of measuring range, additional 0.3 Pa intrinsic error • = 0.02 % of measuring range per degree Kelvin of K slope drift deviation from nominal temperature 22 °C •...
  • Page 10 1 hPa 2500 hPa Upon delivery and following a factory reset the readings are shown in the display in the unit that was ordered via the KMAT option Fxx, see 6.2.1. Ordering options for 6381 transmitter (0555 6381). Meas. cycle •...
  • Page 11 4 Transmitter Maximal load • 4-wire: 10 kΩ (voltage output) Analog output • 0 to 1 V ± 1.5 mV (4-wire) or • 0 to 5 V ± 7.5 mV (4-wire) or • 0 to 10 V ± 15 mV (4-wire) or •...
  • Page 12: Dimensions

    4 Transmitter Housing, weight • Metal, 1.960 kg • Optional Ethernet module: 0.610 kg Protection class • IP 65 only if the transmitter is wired and/or sealing plugs are inserted Directives, standards and tests • EC Directive: 2014/30/EU Pos : 25 /TD/Leistungs bes chr eibung/T ec hnische D aten/MUF 63xx/MUF 63xx Abmes sungen @ 3\mod_1234450671494_79. doc x @ 25194 @ 3 @ 1 4.1.5.
  • Page 13: Product Description

    4 Transmitter 4.2. Product description Pos : 27 /TD/Produkt bes chr eibung/Ü bersicht/MUF 63xx/ Auf einen Blick 6387 @ 4\m od_1252418613327_79.docx @ 48705 @ 3 @ 1 4.2.1. At a glance 1 Keys (only with optional display) 2 Service flap screw connection (self-locking, 2 pcs.) 3...
  • Page 14 4 Transmitter 15 Housing screws 16 Hole for fastening to rear panel bracket (M3 x 6 screw) 17 Plastic bracket for assembly on rear panel Pos : 28 /TD/Produkt bes chr eibung/Ü bersicht/MUF 63xx/Dis play und T ast atur @ 3\mod_1234773965059_79. doc x @ 25650 @ 3 @ 1...
  • Page 15: Display And Keypad

    4 Transmitter 4.2.2. Display and keypad The display option allows operation of the testo 6381 transmitter via the display and four keys. The LCD display consists of two 7-segment lines for displaying readings and units and of an information line (for status messages, for example).
  • Page 16: Parameters

    4 Transmitter 4.2.6. Parameters The following parameters are displayed: • Differential pressure in Pa, hPa, kPa, mbar, bar, mmH O, inch O, inch HG, kg/cm , PSI • Flow in m/s, ft/min • Volumetric flow rate in m /h, l/min, Nm /h, Nl/min •...
  • Page 17 4 Transmitter Pos : 35 /TD/Produkt bes chr eibung/Ü bersicht/MUF 63xx/T abell e Skali erung MUF 638x @ 3\m od_1236343950039_79.docx @ 27502 @ @ 1 Measuring Maximum scaling range/standard scaling 0 to 50 Pa -5 to 15 Pa 0 to 50 Pa -25 to 75 Pa 0 to 100 Pa -50 to 150 Pa...
  • Page 18: Alarm Handling

    The testo 6381 monitors limit values with the help of relays. If a reading is outside the limit values, a relay to be specified by the user is switched.
  • Page 19: Commissioning

    4 Transmitter 4.3. Commissioning Pos : 38 /TD/Erst e Schrit te/MUF 63x x/Wandmont age 635x_6321 @ 3\mod_1236065568006_79.docx @ 27063 @ 3455 @ 1 4.3.1. Assembling the instrument 4.3.1.1. Wall mounting Attaching rear panel bracket 1. Remove locking screw (see item (4) of drawing below) and detach rear panel bracket from plastic bracket (see item (2) of drawing below).
  • Page 20: Connecting The Instrument

    4 Transmitter Pos : 39 /TD/Erst e Schrit te/MUF 63x x/G erät anschli eßen @ 3\m od_1234779466958_79.docx @ 25878 @ 35 @ 1 4.3.2. Connecting the instrument Opening the instrument 1. Loosen screw connection (1) of service flap and open the flap.
  • Page 21 4 Transmitter 2. Loosen and remove housing screws (2). 3. Remove upper part of housing from lower part (3) and place on a clean surface. WARNING Electrical voltage Danger of injury! > De-energize the mains connection before connecting the transmitter. Only have the transmitter wired and connected by authorized personnel with the voltage disconnected.
  • Page 22: Overview Of Terminals

    4 Transmitter Pos : 40 /TD/Erst e Schrit te/MUF 63x x/Ansc hl uss übersicht 638x @ 3\m od_1234780656658_79.docx @ 25935 @ 4 @ 1 4.3.2.1. Overview of terminals 1 Lower part of housing 7 Earthing terminal (internal) 2 Relay board (option) 8...
  • Page 23: Connecting Voltage Supply And Analog Outputs

    4 Transmitter Pos : 41 /TD/Erst e Schrit te/MUF 63x x/Spannungsv ersorg ung/ Analog ausgäng e ans chli eßen 6387/6388 @ 4\mod_1250675956913_79. doc x @ 47503 @ 45 @ 1 4.3.2.2. Connecting voltage supply and analog outputs Terminal strip for voltage supply and analog outputs, item (5) of overview of terminals...
  • Page 24: Connecting The Relay Outputs

    4 Transmitter Requirement for the connecting cable of the supply: • Insulated with cross-section of at least 0.25 mm², maximum 2.7 mm² without wire end sleeves. • The supply line must be secured against exceeding 8 A. • An OFF switch must be installed in an easily accessible position close by and be marked as such.
  • Page 25 4 Transmitter Using plug-in connections (optional) Only insert or disconnect the plug-in connection when the voltage is disconnected. 4. Clean the connector of the probe line and the coupling of any foreign matter. Do not disconnect the connector of the probe line from the instrument for extended periods to protect against contamination.
  • Page 26 4 Transmitter Use of relay as NC contact (NC = normally closed) 1 Alarm/status light (example of installation) 2 250 V AC/DC, 3 A The busy light (alarm/status light) is permanently on until the relay opens or the circuit is interrupted. This circuit can therefore be used to monitor the functionality of the alarm circuit, as a cable break, for instance, is indicated by the busy light going off.
  • Page 27: Plug-In Connection Option

    4 Transmitter The busy light (alarm/status light) only comes on when the relay is switched (closed). Monitoring the functionality of the alarm circuit is therefore not possible with this switching operation. 5.  C lose M 20 x 1.5 screw connection (item (9,) in overview of terminals).
  • Page 28: Creating The Pe/Earthing Terminal

    Pos : 44 /TD/Erst e Schrit te/MUF 63x x/PE-/ Erdungs anschl uss herstellen 638x @ 3\m od_1234788670248_79. docx @ 26049 @ 455 @ 1 4.3.2.5. Creating the PE/earthing terminal As the testo 6381 has a metal housing, we recommend that the instrument be earthed. This can be done using the earthing terminal within the instrument (1) or the earthing terminal outside of the instrument (2).
  • Page 29: Closing The Instrument

    4 Transmitter Pos : 45 /TD/Erst e Schrit te/MUF 63x x/G erät sc hließ en @ 3\mod_1234788511529_79. docx @ 26030 @ 4455 @ 1 4.3.2.6. Closing the instrument 1. Place upper part of instrument on top of lower part (see arrow) and fix in place with housing screws (1).
  • Page 30: Adjusting The Instrument

    4 Transmitter 2. Close the service flap and tighten screws (2). Pos : 46 /TD/Ü berschriften/MUF /5.3.3 Gerät abgleic hen @ 3\m od_1236081099337_79.docx @ 27173 @ 43 @ 1 4.3.3. Adjusting the instrument Pos : 47 /TD/Erst e Schrit te/MUF 63x x/Analog ausg angs-Abgleic h 635x @ 3\m od_1237373284088_79.docx @ 29825 @ 435 @ 1 4.3.3.1.
  • Page 31 3. Contact ch. 1- 4. Service interface Adjusting analog output 1 ✓ With testo 6381 with current output: Load of max. 500 is connected to channel 1 (see 4.3.2.4. Plug-in connections for power supply and channels) ✓ A precise multimeter (minimum requirement: resolution 6.5 digits, at least 5-times more accurate than the 6381) is available.
  • Page 32: N-Point Adjustment

    ✓ A precise pressure sensor (5-times more accurate than the transmitter, e.g. DPC precision pressure sensor from testo industrial services) is available. 1. Connect the positive output of the pressure sensor (3) to the...
  • Page 33: Operation

    4.4.1. Relationship between user menu and mini-DIN socket is active The testo 6381 can be parameterized using either the user menu or the P2A software (see 5. Parameterizing, adjusting and analyzing software (P2A software)). The testo 6381 transmitter can only be operated via the display and keypad if the display option is available.
  • Page 34: Key Cover

    4 Transmitter 4.4.2. Key cover To prevent unauthorized operation of the keys, the standard key frame can be replaced with a key cover. If the key cover has been assembled, the service flap must be opened for operation. Attaching the key cover ✓...
  • Page 35: Structure Of User Menu

    4 Transmitter Pos : 53 /TD/Produkt verw enden/MUF 63xx /Auf bau des Bedi enm enüs 635x @ 3\mod_1236162940119_79.doc x @ 27333 @ 33 @ 1 4.4.4. Structure of user menu At the main menu level, the user menu comprises the following: •...
  • Page 36: Overview Of The Testo 6381 User Menu

    4 Transmitter Pos : 54 /TD/Produkt verw enden/MUF 63xx /Übersic ht Bedi enm enü @ 3\m od_1234510821302_79.docx @ 25303 @ 33 @ 1 4.4.5. Overview of the testo 6381 user menu...
  • Page 37 4 Transmitter...
  • Page 38 4 Transmitter...
  • Page 39: The Individual Main Menus

    The individual main menus 4.4.6.1. Editing main menu of channel 1 An overview is given in 4.4.5. Overview of the testo 6381 user menu). You can perform basic settings for channel 1. 1. In the Measuring Mode press SET, select...
  • Page 40: Editing Main Menu Alarm

    4 Transmitter 4.4.6.2. Editing Main Menu Alarm With the alarm, the relays, available as options, are programmed. In addition, the alarm statuses are shown on the display (top right) (even without relays). You can choose whether the alarm is to be used to monitor limit values or as a collective alarm.
  • Page 41 If an alarm is assigned to the collective alarm, the relay is switched and a visual alarm can be issued via the display as soon as (at least) one of the warning or error messages of the testo 6381 transmitter becomes active.
  • Page 42: Editing Main Menu Settings

    4 Transmitter 4.4.6.3. Editing Main Menu Settings You can edit instrument settings and other settings. > In Measuring Mode, press SET, select Main Menu Settings using  or  and confirm selection with SET. You can edit settings for: • Display •...
  • Page 43 4 Transmitter Selecting language You can select the language for the plain text line in the display. > Press SET, select required language with  or , confirm selection with and return to Language. Only choose a language that you can understand well. Select unit This setting affects the unit of the standard and process data.
  • Page 44: Editing Main Menu Analysis

    ESC. 3. Accept setting by pressing and test with multimeter (minimum requirement: resolution 6.5 digits, at least 2-times more accurate than the 6381) as follows: Analog output 1: Via test contacts under service flap, see diagram.
  • Page 45 4 Transmitter 1 Channel 1 test contacts 2 Service interface 3 Multimeter 4. Return to Test Analog Output with and use or  to continue to Test Relay Output. Testing functionality of the pressure sensor (Test pressure sensor) This function is only required to calibrate the pressure sensor.
  • Page 46: Editing Message Main Menu

    4 Transmitter 4.4.6.5. Editing Message main menu Messages can be confirmed/acknowledged, the last messages can be called up and the display of the messages can be switched on or off. 1 Operating hours at the time of message 2 Message code (see 4.5. Status, warning and error messages).
  • Page 47: Calling Up Main Menu Ident

    4 Transmitter An overview of the messages can be found in 4.5. Status, warning and error messages. Pos : 60 /TD/Produkt verw enden/MUF 63xx /Hauptm enü Ident abfragen 635x @ 3\m od_1238158282021_79.docx @ 30233 @ 44 @ 1 4.4.6.6. Calling up Main Menu Ident 1...
  • Page 48 4 Transmitter 5. Use  or  to select Adj. Point 6. Press SET. Read off multimeter display and enter this value in the user menu. Do this by scrolling one digit to the right using  and increasing the value of digit by 1 using . Confirm with or abort entry with ESC.
  • Page 49: Editing Reset Main Menu

    Warning messages • Error messages in each case for the testo 6381 transmitter. All messages are stored in the transmitter with an operating hours stamp. Use the user menu (see 4.4.6.5. Editing Message main menu) or the P2A software (see 5.3.5. Transmitter history) to view the message history.
  • Page 50: Status Messages

    4 Transmitter 4.5.1. Status messages Status messages show the current operating status of the testo 6381. Message Display Description 02506 Sensor Message appears while the initialization transmitter is starting up. If the message disappears, the transmitter is ready for operation.
  • Page 51: Warning Messages

    4 Transmitter 4.5.2. Warning messages Warning messages show an early warning or a current malfunction which may negatively impact measuring. Message Display Cause Remedying of fault 00E00 T ambient high The ambient Take necessary temperature exceeds measures to lower the permissible ambient temperature, temperature for the e.g.
  • Page 52: Transmitter Error Messages

    Cause Remedying of fault 01505 Watchdog error Due to a processor If the problem occurs error, the transmitter frequently, contact performs an automatic Testo Service. restart. 01115 Low adjustment The ambient Take necessary temperature temperature is too low measures to raise...
  • Page 53: Namur Fault Conditions

    4 Transmitter Shown on the Can be used for Additional display collective alarm message end Analog adjustment T ambient high T ambient low Supply voltage low Watchdog error Perform the Confirm message function (acknowledgement of the alarm via the control keys on the transmitter): •...
  • Page 54: Maintenance And Cleaning

    4 Transmitter Pos : 70 /TD/Ü berschriften/MUF /4.6 W artung und Rei nigung @ 3\m od_1234443039129_79. docx @ 24982 @ 23 @ 1 4.6. Maintenance and cleaning Pos : 71 /TD/Produkt i nstand halten/MUF 63x x/Gerät w arten/r einigen 635x @ 4\m od_1245756987352_79.docx @ 45305 @ 233 @ 1 4.6.1.
  • Page 55: Parameterizing, Adjusting And Analyzing Software (P2A Software)

    Pos : 74 /TD/Leistungs bes chr eibung/Verwendung/MUF63xx/MUF 63xx P2A @ 3\mod_1234258967326_79.docx @ 24065 @ 3255 @ 1 The P2A software is used for the parameterizing, adjustment and analysis of testo transmitters. The following applies: • Generally, all newer testo transmitters (as of 2007) are supported. • For each newly purchased Testo transmitter, a free software upgrade must be installed, containing the instrument drivers for all transmitters which are connectable at that time.
  • Page 56: System Requirements

    5 Parameterizing, adjusting and analyzing software (P2A software) Parameter file Parameter files are not tied to a specific individual transmitter and contain only parameter data/no history data. If you use various instruments of the same type, you can create parameter files once (e.g. by saving the appropriate instrument file as the parameter file) and transmit these onto the other instruments.
  • Page 57: Scope Of Delivery

    Without the input of a licence key, the software will run only as a demo version (time limit 30 days). 1. You can download the software under the following link: https://www.testo.com/download-center. If the installation program does not start automatically: > Open download folder and start P2A.exe.
  • Page 58: Starting The Software

    Without the input of a licence key, the software will run only as a demo version (time limit 30 days). 4. You can download the software under the following link: https://www.testo.com/download-center. If the installation program does not start automatically: > Open download folder and start P2A.exe.
  • Page 59: Using The Software

    5 Parameterizing, adjusting and analyzing software (P2A software) If a connection with the instrument is established when the program is started, the corresponding instrument file is marked automatically. Pos : 80 /TD/Ü berschriften/MUF /6.3 Softw are verw enden @ 3\m od_1234258679599_79.docx @ 23989 @ 23444 @ 1 5.3.
  • Page 60 5 Parameterizing, adjusting and analyzing software (P2A software) Menu Command Explanation Check instrument Checks the connections to a connections connected instrument without the instrument having to be activated. Service A text file with the most important information on the computer and the software is opened via Display service data.
  • Page 61: Editing Instrument/Parameter File

    5 Parameterizing, adjusting and analyzing software (P2A software) [Change parameterization] see 5.3.2.1. Changing instrument/parameter file. [Test/analyze transmitter] see 5.3.3. Chapter Analyzing/testing the transmitter. [Adjusting the transmitter] see 5.3.4. Chapter Adjusting the transmitter. [Transmitter history] see 5.3.5. Transmitter history. 5 File information: Status Shown in the window An instrument file is...
  • Page 62 5 Parameterizing, adjusting and analyzing software (P2A software) 2. Change or enter parameters in the corresponding fields. Pos : 83 /TD/Produkt verw enden/MUF 63xx /P2A/Ei nheit / Anal ogausgang @ 3\m od_1234366217517_79.docx @ 24592 @ 34 @ 1 Field Explanation Unit/ All analog outputs are parameterized in this mask.
  • Page 63 5 Parameterizing, adjusting and analyzing software (P2A software) Field Explanation Unit Selection of the physical unit. When changing the unit, standard values are set for scale minimum and maximum. Caution! When changing the phys. unit, the relay limit values are set to the assigned default values. Signal delay Curve changes according to the set signal (graphic)
  • Page 64 5 Parameterizing, adjusting and analyzing software (P2A software) Pos : 84 /TD/Produkt verw enden/MUF 63xx /P2A/Gr enz wert e R elais @ 3\mod_1234366741293_79. doc x @ 24611 @ @ 4 Field Explanation Limit values, relay 1 to 4/alarm In this mask, the relays or display values, alarm 1 to 4 alarms are parameterized.
  • Page 65 5 Parameterizing, adjusting and analyzing software (P2A software) Field Explanation Max control If switched to ON (NO contact) or OFF (NC contact) above the limit value; in the event of a subsequent undershooting of Limit value minus Hysteresis, it is switched to OFF (NO contact) or ON (NC contact).
  • Page 66 (0 to 3600 seconds possible). The alarm delay has no effect on the collective alarms. Pos : 85 /TD/Produkt verw enden/MUF 63xx /P2A/Pr oz ess daten Druck+Norm daten 635x /6381 @ 3\m od_1235994444633_79.docx @ 26986 @ @ 1 Field Explanation Basic settings...
  • Page 67 5 Parameterizing, adjusting and analyzing software (P2A software) Field Explanation Pitot tube factor The Pitot tube correction factor depends on the shape of the Pitot tube (see operating instructions of the Pitot tube). The entered correction factor has a direct proportional effect on the volumetric flow rate.
  • Page 68 5 Parameterizing, adjusting and analyzing software (P2A software) Pos : 86 /TD/Produkt verw enden/MUF 63xx /P2A/Displ ay @ 3\m od_1234424330866_79. docx @ 24673 @ @ 1 Field Explanation Display Setting the display functions (if a display is available on the transmitter). Continuous display Display lighting is permanently switched on.
  • Page 69: Saving Parameters

    5 Parameterizing, adjusting and analyzing software (P2A software) Field Explanation New password The password consists of four numbers, each of which must be between 1 and 9. If the password protection is not to be used, the numerical code "0000" must be entered. [Adopt new Button for confirming the new password.
  • Page 70: Opening The Parameter File

    The original name (Instrument type, Serial number) is suggested with the current date/time as standard, e.g. "testo 6381 01234578 061120 1403.cfp". For a standard installation, the files are saved under "C:\Documents and Settings\All Users\Shared Documents\P2A Software". The path can differ depending on the version of the operating system.
  • Page 71: Deleting Instrument/Parameter File

    5 Parameterizing, adjusting and analyzing software (P2A software) Pos : 91 /TD/Produkt verw enden/MUF 63xx /P2A/G eräte-/ Param eter datei lösc hen @ 3\m od_1234359025194_79.docx @ 24379 @ 44 @ 1 5.3.2.5. Deleting instrument/parameter file Instrument/parameter files can be deleted from the file list. 1....
  • Page 72: Carrying Out Factory Reset

    5 Parameterizing, adjusting and analyzing software (P2A software) Pos : 95 /TD/Produkt verw enden/MUF 63xx /P2A/Werksr es et durchführen @ 3\mod_1237379864072_79. docx @ 29887 @ 44 @ 1 5.3.3.2. Carrying out factory reset ✓ The required instrument file is marked. 1....
  • Page 73: Testing Switch Output Relays 1 To 4

    5 Parameterizing, adjusting and analyzing software (P2A software) Field/button Explanation Unit Unit according to the type of analog output. Default value Freely definable output value for the respective type of analog output (V or mA), 1 decimal place. [Activate] The entered default value is forwarded to the corresponding analog output and to the test contacts by clicking.
  • Page 74: Displaying Min./Max. Values

    5 Parameterizing, adjusting and analyzing software (P2A software) 2. Mark Relay tests and test the values. Field/button Explanation Test the relay function (see 4.4.6.4. Testing functionality of relay outputs). [Activate relay n] Close contact. A warning informs that the value is being transmitted to a connected PLC, external display, etc.
  • Page 75: Adjusting The Transmitter

    5 Parameterizing, adjusting and analyzing software (P2A software) 2. Mark Min./max. values. Field/button Explanation Min./max. values View the min./max. values of each channel. Only the values within the measuring range are shown. Channel Channel 1 min./max. Value Min. or max. value, 1 decimal place. Unit Unit selected in Unit/analog...
  • Page 76: N-Point Adjustment

    5 Parameterizing, adjusting and analyzing software (P2A software) Pos : 101 /TD/ Produk t v erwenden/MUF 63xx/ P2A/ n-Punkt-Abgleic h _635x @ 3\m od_1237457345714_79. docx @ 30083 @ 4 @ 1 5.3.4.1. n-point adjustment 1. Connect precision pressure sensor (see 4.3.3.2. n-point adjustment).
  • Page 77: Adjusting The Analog Output

    5 Parameterizing, adjusting and analyzing software (P2A software) Pos : 102 /TD/ Produk t v erwenden/MUF 63xx/ P2A/ Anal ogausgang abgleic hen _635x @ 3\m od_1234363369967_79.docx @ 24531 @ 44 @ 1 5.3.4.2. Adjusting the analog output 1. Connect precision multimeter (see 4.3.3.1. Analog output adjustment).
  • Page 78: Transmitter History

    5 Parameterizing, adjusting and analyzing software (P2A software) Field Explanation Measured analog Required field: Entry of the value read off at value the multimeter. Pos : 103 /TD/Ü bersc hrift en/MUF/ 4.3. 5 M ess umform er-Histori e @ 3\mod_1237373806926_79.doc x @ 29846 @ 43 @ 1 5.3.5.
  • Page 79 5 Parameterizing, adjusting and analyzing software (P2A software) Field Explanation Operating hours / Operating hour/time stamp at which the date/time change at the instrument was performed. User Name with which the user is logged into the operating system. "Transmitter" entry if the change was performed at the instrument.
  • Page 80 5 Parameterizing, adjusting and analyzing software (P2A software) Field Explanation Selection: Analog adjustments. Operating hours / Operating hour/time stamp at which the date/time change at the instrument was performed. Name with which the user is logged into the User operating system. "Transmitter"...
  • Page 81 5 Parameterizing, adjusting and analyzing software (P2A software) Field Explanation The table is shown only for error and status messages that were generated in the transmitter and were transferred and saved there via the connection to the P2A software. Operating hour Operating hour at which the message appeared in the instrument.
  • Page 82: Tips And Assistance

    If we could not answer your question, please contact your dealer or Testo Customer Service. For contact details see the rear side of this document or the web page www.testo.com/service-contact Pos : 108 /TD/Ü bersc hrift en/8.3 Zubehör und Ersatzteile @ 0\m od_1177402058734_79.docx @ 1102 @ 2 @ 1 6.2.
  • Page 83: Ordering Options For 6381 Transmitter (0555 6381)

    Pos : 110 /TD/Ti pps und Hilfe/Z ubehör und Ers atzteile/MUF63xx /Bestellopti onen MUF 6380 @ 3\mod_1234277777760_79.docx @ 24261 @ 3 @ 1 6.2.1. Ordering options for 6381 transmitter (0555 6381) Order code Characteristic Axx Measuring range...
  • Page 84 6 Tips and assistance Order code Characteristic -10 to 10 hPa -50 to 50 hPa -100 to 100 hPa -500 to 500 hPa -1000 to 1000 hPa Bxx Analog output/supply 0 to 1 V (4-wire, 24 V AC/DC) 0 to 5 V (4-wire, 24 V AC/DC) 0 to 10 V (4-wire, 24 V AC/DC) 0 to 20 mA (4-wire, 24 V AC/DC) 4 to 20 mA (4-wire, 24 V AC/DC)
  • Page 85 /Min/Max ft/min /Min/Max /h /Min/Max l/min /Min/Max /min /Min/Max Nl/min /Min/Max Gxx Optional analog output for humidity probe connection testo 6610/units Without connection option for testo 6610 humidity probe % RH/Min/Max °C/Min/Max °F/Min/Max °C /Min/Max °F /Min/Max g/kg /Min/Max...
  • Page 86 4 relay outputs, limit value monitoring 4 relay outputs, channel 1 limit values and collective alarm Ixx Units, channel 3 (only if optional humidity probe connection is available) Without testo 6610 humidity probe % RH/Min/Max °C/Min/Max °F/Min/Max °C /Min/Max °F /Min/Max...
  • Page 87 6 Tips and assistance Order code Characteristic /Min/Max gr/ft /Min/Max ppm (vol)/Min/Max °C /Min/Max °F /Min/Max kJ/kg /Min/Max (enthalpy) mbar/Min/Max (water vapour partial pressure) inch H O/Min/Max (water vapour partial pressure) °C °F % vol === Ende der Liste f ür T extm arke I nhalt ===...
  • Page 88 6 Tips and assistance...
  • Page 89 6 Tips and assistance...
  • Page 90 6 Tips and assistance...
  • Page 92 0970 6387 en 01 V01.0 V01.40-1 en-GB...

Table of Contents