TESTO 6381 Instruction Manual page 141

Differential pressure transmitter, probes
Hide thumbs Also See for 6381:
Table of Contents

Advertisement

7 Parameterizing, adjusting and analyzing software (P2A software)
Field
Signal delay
The delay of the signal in relation to the change in the
process is also significantly influenced by the selection of
the particle filter.
Field
Limit values, relay 1 to 4/alarm
values, alarm 1 to 4
Relay x/alarm x
Not used
Explanation
Time interval in stages 1 – 15:
1 = no delay
15 = longest delay.
The signal delay is added to the reaction time
of the sensor. The signal delay shows
averaging over the time interval of the
selected stage in seconds:
Example
Stage 10 = average of the readings from the
last 10 seconds.
Explanation
In this mask, the relays or display
alarms are parameterized.
Four relays or alarm values are
available (optional).
Relay is not used.
Hysteresis image and input
options are hidden.
141