Interface
The communication takes place according to the client/server
principle, whereby the Ethernet module takes on the role of the
server:
The XML interface will be approached via a URL. The URL consists
of the IP address of the Ethernet module, the path for the XML
document and, depending on the URL, with a clearly defined
parameter. If a parameter is missing or an incorrect value is
transferred, the Ethernet module sends back a simple error
message.
Example:
IP address = 254.169.100.100
URL for serialnumber.xml:
http://254.169.100.100/data/getserialnumber
If a parameter is required, a "?" is always in the first place after the
URL, then the parameter is transferred to the common query string
form (name=value).
Example:
IP address = 254.169.100.100
URL for / identification.xml:
http://254.169.100.100/data/getidentification?param=0 (compare
table xml codes)
After calling up the compound URL, if everything was correct, an
XML document is returned. With faulty parameters a simple
HTML answer comes that points out the reason that an XML
answer could not be given.
If an XML document is to be loaded on the server, access is
provided by means of a POST request
Example of an upload via the program wget:
IP address = 254.169.100.100
URL for / usersettings.xml:
Path details of the folder in which
the wget program is located
H:/wget/wget-complete-stable
The Ethernet module supports reading out of
Readings
Instrument type (testo 6381)
Firmware date and version (testo 6381)
Status and status messages (testo 6381)
Path details of the
Command
folder in which the
usersettings XML file is
located
/wget --post-file=
C:/usersettings.xml
4 Transmitter
IP address of the transmitter
and required URL (see table on
the next page)
254.169.100.100/config/
setusersettings
41