Technical data
Parameters
•
Humidity (% RH/°C
•
Temperature
Measuring range
•
Humidity: 0 to 100 % RH
•
Temperature: -20 to +70 °C/-4 to +158 °C
Material
•
Probe shaft: Stainless steel
•
Connector: ABS plastic
Accuracy (at 25 °C/77 °F)*
•
Length 200 mm
•
Humidity
◦
◦
11
For constant use in high humidity (> 80% rF at ≤ 30°C for > 12 h, > 60% rF at > 30°C for > 12 h), please contact
us via www.testo.com/service-contact
/°F
td
± (1.0 % RH + 0.007 x reading) for 0 to 90 % RH
± (1.4 % RH + 0.007 x reading) for 90 to 100 % RH
) etc.
td
11
6 testo 6610 probes
105