4 Transmitter
7. Edit humidity process pressure: Scroll one digit to the right
using  and increase value of digit by 1 using . Confirm with
SET
8. Return to
9. Return to
4.4.6.6.

Editing Main Menu Analysis

You can test the functionality of analog and relay outputs. In
addition, you can read off the minimum and maximum values (since
the last voltage supply or reset of the min./max. values).
Testing functionality of analog outputs
1. In the Measuring Mode press SET, select main menu
with  or  and confirm selection with SET.
-
Test Analog Output
2. Press
or .
3.  P ress SET, scroll one digit to the right using  and increase
value of digit by 1 using . Any analog output value can be
predefined, e.g. for an analog output of 4 to 20 mA, the value
"6.0 mA". Confirm with
4. Accept setting by pressing
(minimum requirement: resolution 6.5 digits, at least 2-times
more accurate than the 6381) as follows:
Analog output 1 or 2: Via test contacts under service flap, see
diagram.
Analog output 3: Connect measuring cables to terminals for
channel 3, guide measuring cables out of housing and perform
measurement outside of the transmitter, see diagram.
74
or abort entry with ESC.
Change parameters
Main Menu Settings
This function affects the analog outputs directly, not only
the test contacts.
is shown.
SET
and choose between
with ESC.
with ESC.
Analog output 1, 2, 3
SET
or abort entry with ESC.
SET
and test with multimeter
Analysis
with 
1 Channel 1
test contacts
2 Service
interface
3 Channel 2
test contacts
4 Ends of
channel 3
cables
5 Multimeter