Adjusting The Analog Output - TESTO 6381 Instruction Manual

Differential pressure transmitter, probes
Hide thumbs Also See for 6381:
Table of Contents

Advertisement

7 Parameterizing, adjusting and analyzing software (P2A software)
3. Click on
-
The
dialogue is opened with the
4. Mark
5. Click on
wizard.
The adjustment is performed when the wizard is closed.
-
Field
How much pressure
is actually applied
7.3.4.4.

Adjusting the analog output

1. Connect precision multimeter (see volume 1, 4.3.4.10. Analog
output adjustment).
2. Mark the instrument file of the connected instrument.
[Adjusting the
Properties of <Instrument type> <Serial number>
n-point
adjustment.
[Start wizard ...]
Explanation
Required field: Entry of the value read off at
the pressure sensor.
The n-point adjustment must always be carried out to its full
extent and in good time at all selected adjustment points.
The number of adjustment points (3 to 6) is stored in the
user menu of the transmitter and can only be changed via
the P2A software.
transmitter].
Adjusting the transmitter
and follow the instructions of the
register.