6 testo 6610 probes
Technical data
Parameters
•
Humidity
•
Temperature
Measuring range
•
Humidity: 0 to 100 % RH
•
Temperature: -70 to +180 °C/-94 to +356 °F
Material
•
Probe shaft: Stainless steel
•
Line: Sheathed, FEP
•
Connector: ABS plastic
Accuracy (at 25 °C/77 °F)*
•
Humidity
◦
◦
◦
◦
For constant use in high humidity (> 80% rF at ≤ 30°C for > 12 h, > 60% rF at > 30°C for > 12 h), please contact
13
us via www.testo.com/service-contact
112
± (1.0 % RH + 0.007 x reading) for 0 to 90 % RH
± (1.4 % RH + 0.007 x reading) for 90 to 100 % RH
0.02 % RH/K, dependent upon the process temperature
(with a deviation of 25 °C/+77 °F)
0.02 % RH/K, dependent upon the electronics temperature
(with a deviation of 25 °C/+77 °F)
13