screw (3), washer (4) and snap ring (7) on the internal earthing
terminal (1).
2. Place the other cable end on an appropriate functional earth,
e.g. an earthing bar.
Using an earthing terminal outside of the instrument
1. Use PE line (yellow-green) (5) with cable lug (8). Fix this using
M 5 screw (3), washer (4) and snap ring (7) on the external
earthing terminal (2).
2. Place the other cable end on an appropriate functional earth,
e.g. an earthing bar.
4.3.3.6.

Setting the Ethernet module

Using a DIP switch the Ethernet module can be switched between
two main functions:
Saveris subscriber function (DIP switch no. 1 = on, DIP switch
no. 2 = on), testo 6381 is used as a testo Saveris component.
XML server function (DIP switch no. 1 = off, DIP switch no. 2 =
on), integration of the testo 6381 into the customer's Ethernet
system.
✓ The testo 6381 must be separated from the Ethernet network.
1. Open transmitter (see 4.3.3. Opening the instrument).
In the condition on delivery, the switch position is set to on
(Saveris mode) with the testo 6381 with Ethernet module.