TESTO 6381 Instruction Manual page 61

Differential pressure transmitter, probes
Hide thumbs Also See for 6381:
Table of Contents

Advertisement

All adjustment processes are saved in the history of the transmitter,
see volume 2, 7.3.5. Transmitter history.
Deactivating the adjustment function of the testo 6615
will reduce measuring accuracy and should therefore be
restricted to the shortest possible length of time.
During the heating phase, the relay and analog outputs,
the display value and output value are "frozen", see
diagram above.
display until it has finished. The factory setting for self-
adjustment time (incl. heating time, calculation time,
cooling time) requires 30 minutes daily. The cycle time
can be edited via the operating menu or the P2A
software.
In the factory setting, a third adjustment point
(-40 °C
the 2-point adjustment. This special adjustment can be
performed again by your Testo Service team if
necessary.
Self-adjustment active
) is approached for the testo 6615 in addition to
td
4 Transmitter
is shown in the