6. Return to
continue to Test Relay Output.
Testing functionality of the pressure sensor
sensor)
Testing functionality of relay outputs
1. Press
2. Press SET.
The relay can now be tested. You can choose between
and
closed, the NC contact opened. If
contact is closed, the NO contact opened.
3. To test, route a measuring cable from the relay terminals (see
4.3.3.3. Connecting the relay outputs) out of the transmitter to a
multimeter (resistance measurement) or continuity tester.
4. Return to
ESC
Reading off min./max. values of channels
To reset the max./min. values, see 4.4.6.10. Editing Reset main
menu.
1. Read off the min./max. values of the three channels in
succession with or and return to the Main Menu Analysis
with ESC.
2. Continue to
Measuring Mode with ESC.
4.4.6.7.
Editing Message main menu
Messages can be confirmed/acknowledged, the last messages can
be called up and the display of the messages can be switched on
or off.
Test Analog Output
This function is only required to calibrate the pressure
sensor.
SET
and choose between
ON
using or . If
Test Relay Output
(exits the menu without relay test).
Main Menu Message
and use or to
with
ESC
Alarm 1, 2, 3, 4
ON
is chosen, the NO contact is
OFF
is chosen, the NC
with
SET
(starts relay test) or
with or or return to
4 Transmitter
(Test pressure
with or .
OFF
75