7 Parameterizing, adjusting and analyzing software (P2A software)
7.3.4.
Adjusting the transmitter
This function is used to adjust an attached instrument. The
following adjustments may be carried out using the software:
•
1-point adjustment (offset)
•
2-point adjustment (upper and lower adjustment point)
•
Analog adjustment (entry via assistant/wizard)
•
n-point adjustment (entry via assistant/wizard)
Also see volume 1, 4.3.4.6. Adjusting the instrument.
7.3.4.1.
1-point adjustment
1. Expose the reference measuring instrument and the instrument
to be adjusted to the same constant conditions and wait for
equalization period to lapse.
2. Mark the instrument file of the connected instrument.
3. Click on
-
The
dialogue is opened with the
4. Enter reference value and click on
adjustment].
5. Confirm confirmation request.
-
The adjustment is carried out.
158
The testo 400/650 with precision humidity probe (order no.
0636 9741) is recommended as the reference measuring
instrument for 1-point adjustment (offset) (see volume 1,
4.3.4.8. Adjusting testo 6381 using testo handheld
instrument).
[Adjusting the
Properties of <Instrument type> <Serial number>
transmitter].
Adjusting the transmitter
[Carry out 1-point
register.