4 Transmitter
Enter reference value for 1-point adjustment
Please also refer to 4.3.4.8. 1-point adjustment (offset -
humidity/temperature)
1. In the Measuring Mode press SET, select main menu
with  or  and confirm selection with SET.
-
Reference value % RH
2. Press SET, edit value: Scroll one digit to the right using  and
increase value of digit by 1 using . Confirm with
entry with ESC.
3. Continue to
4. Press SET,
displayed.
5. Press SET, edit value: Scroll one digit to the right using  and
increase value of digit by 1 using . Confirm with
entry with ESC.
6. Continue to
7. Continue with the adjustment of the analog outputs (see below,
step 2) or press
8. Continue to main menu
Measuring Mode with ESC.
Performing the 2-Points-Adjustment for 20% RH and 80 % RH
1. In the Measuring Mode press SET, select main menu
with  or  and confirm selection with SET.
2. Continue to
3. Press SET.
4. Continue to
5. Confirm selection with
-
The red LED
-
In the display
reset
-
The adjusted measurement values appear after probe reset
78
The 2-point adjustment for the adjustment points 20 %RH
and 80 %RH is carried out via the user menu.
For the adjustment points 11.3 %RH and 75.3 %RH, the 2-
point adjustment is carried out via the adjustment buttons or
the P2A software, see 4.3.4.9. 2-point adjustment
(humidity/temperature), or volume 2, 7.3.4.2. 2-point
adjustment.
Reference value temp
Reference value °C
Analog Adj. Ch. 1
ESC
to return to main menu Adjust.
2Pt Adjust 20/80%
Adj. Point 20%
ADJ
flashes
2Pt Adjust
appears consecutively.
is displayed.
with  or .
or
Reference value °F
with  or .
with  or  or return to
Reset
with  or .
or
Adj. Point 80%
SET
20/80%,
1-point
Adjust
SET
or abort
is
SET
or abort
Adjust
with  or .
adjustment,
Probe