4 Transmitter
Enter reference value for 1-point adjustment
Please also refer to 4.3.4.8. 1-point adjustment (offset -
humidity/temperature)
1. In the Measuring Mode press SET, select main menu
with or and confirm selection with SET.
-
Reference value % RH
2. Press SET, edit value: Scroll one digit to the right using and
increase value of digit by 1 using . Confirm with
entry with ESC.
3. Continue to
4. Press SET,
displayed.
5. Press SET, edit value: Scroll one digit to the right using and
increase value of digit by 1 using . Confirm with
entry with ESC.
6. Continue to
7. Continue with the adjustment of the analog outputs (see below,
step 2) or press
8. Continue to main menu
Measuring Mode with ESC.
Performing the 2-Points-Adjustment for 20% RH and 80 % RH
1. In the Measuring Mode press SET, select main menu
with or and confirm selection with SET.
2. Continue to
3. Press SET.
4. Continue to
5. Confirm selection with
-
The red LED
-
In the display
reset
-
The adjusted measurement values appear after probe reset
78
The 2-point adjustment for the adjustment points 20 %RH
and 80 %RH is carried out via the user menu.
For the adjustment points 11.3 %RH and 75.3 %RH, the 2-
point adjustment is carried out via the adjustment buttons or
the P2A software, see 4.3.4.9. 2-point adjustment
(humidity/temperature), or volume 2, 7.3.4.2. 2-point
adjustment.
Reference value temp
Reference value °C
Analog Adj. Ch. 1
ESC
to return to main menu Adjust.
2Pt Adjust 20/80%
Adj. Point 20%
ADJ
flashes
2Pt Adjust
appears consecutively.
is displayed.
with or .
or
Reference value °F
with or .
with or or return to
Reset
with or .
or
Adj. Point 80%
SET
20/80%,
1-point
Adjust
SET
or abort
is
SET
or abort
Adjust
with or .
adjustment,
Probe