7 Parameterizing, adjusting and analyzing software (P2A software)
Field
°C/°F
Current reading
Reference value
> Click on
value.
-
The current reading is used again.
6. Click on
7.3.4.2.
2-point adjustment
Also see volume 1, volume 1, 4.3.4.9. 2-point adjustment
(humidity/temperature).
1. Expose the reference measuring instrument and the instrument
to be adjusted to the same constant conditions and wait for
equalization period to lapse.
2. Mark the instrument file of the connected instrument.
Explanation
Selection of the unit; only for temperature
adjustment.
Reading in °C/°F or % RH.
Readings are updated every second.
Entry of the read-off value from the reference
measuring instrument.
Permissible entries:
•
max. 5 % RH deviation (sum of all 1-point
adjustments)
•
max. 2 K (°C) deviation (sum of all 1-point
adjustments)
[Set offset to zero]
[OK]
or
[Cancel]
to close the dialogue.
to reset a transmitted reference
159