7 Parameterizing, adjusting and analyzing software (P2A software)
3. Click on
-
The
dialogue is opened with the
4. Click on
wizard.
The adjustment is performed when the wizard is closed.
-
Field
Default value
Measured analog
value
162
[Adjusting the
Properties of <Instrument type> <Serial number>
[Start wizard ...]
Explanation
The analog output value from the last
performed adjustment is given at the output.
Value of the factory adjustment:
•
Lower adjustment point: approx. 10 % of
the max. value
•
Centre adjustment point: approx. 50 % of
the max. value
•
Upper adjustment point: approx. 90 % of
the max. value
Required field: Entry of the value read off at
the multimeter.
transmitter].
Adjusting the transmitter
and follow the instructions of the
register.