N-Point Adjustment - TESTO 6381 Instruction Manual

Differential pressure transmitter, probes
Hide thumbs Also See for 6381:
Table of Contents

Advertisement

7 Parameterizing, adjusting and analyzing software (P2A software)
3. Click on
-
The
dialogue is opened with the
4. Mark
point and click on
-
The adjustment is carried out.
5. Mark
adjustment point and click on
The adjustment is carried out.
-
Field
Current reading
Reference value
6. Click on
7.3.4.3.

n-point adjustment

1. Connect precision pressure sensor (see volume 1, 4.3.4.11. n-
point adjustment (pressure)).
2. Mark the instrument file of the connected instrument.
160
[Adjusting the
Properties of <Instrument type> <Serial number>
11.3
%, enter the reference value for the lower adjustment
[Lower adjustment
75.3
%, enter the reference value for the upper
Explanation
Reading in % RH.
Readings are updated every second.
Entry of the read-off value from the reference
measuring instrument.
Permissible entries:
Lower adjustment point 10.3 to 12.3 % RH
Upper adjustment point 74.3 to 76.3 % RH.
[OK]
or
[Cancel]
transmitter].
Adjusting the transmitter
point].
[Upper adjustment
to close the dialogue.
register.
point].